PDGA.100 – Portable Dissolved Gas Analyser
The PDGA.100 Portable Dissolved Gas Analyser is a compact chromatographic system designed for field and laboratory detection of dissolved gases in transformer and insulating oils.
The system directs samples into a dual-column separation setup:
Column 1 (GDX502): Separates hydrocarbon compounds, detected by FID1.
Column 2 (TDX-01): Separates H₂, CO, and CO₂. H₂ is examined via TCD, while CO and CO₂ are transformed into CH₄ through a re-burner and detected by FID2.
This dual-column system ensures high sensitivity and accuracy for dissolved gas analysis, compliant with IEC-60567-2011 and IEC-60599-1999 standards.
Key Features
-
Dual-column chromatographic system for complete gas analysis.
-
Simultaneous detection of H₂, CO, CO₂, CH₄, C₂H₂, C₂H₄, and C₂H₆.
-
Minimum detection limits:
-
H₂: ≤2 ppm
-
CO/CO₂: ≤2 ppm
-
Hydrocarbons: ≤0.1 ppm
-
-
Large 7-inch LCD touch screen for user-friendly operation.
-
Stores up to 100 test results with power-down protection.
-
Built-in thermal printer for real-time reporting.
-
RS232 & USB interface for PC connectivity.
-
Over-voltage and over-current protection for operator safety.
-
Complies with IEC standards for dissolved gas analysis in transformer oil.
Technical Specifications
-
Temperature Control Range:
-
Oven: RT + 6°C to 400°C (1°C step)
-
FID, TCD, Injector, Re-burner: RT + 8°C to 400°C
-
-
Temperature Accuracy:
-
TCD: ±0.02°C
-
Oven, FID, Injector, Re-burner: ±0.1°C
-
-
Detection Limits & Sensitivity:
-
FID Limit of Detection: ≤2×10⁻¹² g/s (N-C16)
-
TCD Sensitivity: ≥8000 mV·mL/mg (N-C16)
-
-
Sample Volume: 5 mL – 50 mL
-
Operation Pressure: 0.08 – 0.3 MPa (800 – 3000 mbar)
-
Operation Temperature: 0°C – 65°C
-
Input Power: 220 VAC / 50 Hz
-
Dimensions: 380 × 500 × 220 mm (L × W × H)
Standard Accessories
-
Gas Extraction Syringe (1)
-
3-Way Valve (1)
-
Plunger Rubber Head (1)
-
Rubber Cap (1)
-
Injection Septum (1)
-
5 mL Glass Syringe for gas collection (1)
-
1 mL Glass Syringe for sample injection (1)
-
200 mL Sampling Container with PU Pipe (1)
-
Nut for Gas Circuit (1)
-
Silicon Rubber Sealing Ring for Gas Circuit (1)
-
WH500-USB Workstation Installation Disk (1)
-
USB Cable (1)
-
Industrial Transport Case (IP67 rated, when closed) (1)
.png)