MTS-100 Meter Test System
The MTS-100 Meter Test System is a three-phase, high-accuracy calibration system designed for the testing and verification of all types of single-phase and three-phase electronic or inductive active and reactive energy meters.
With a built-in reference meter (class 0.05 or 0.1) and a high-precision power source (up to 120A), it provides a complete modular solution for meter manufacturers, research institutes, utilities, and calibration laboratories.
Key Features
-
Class 0.05% or 0.1% accuracy for high reliability in calibration
-
Modular structure: suitable for single/three-phase energy meters
-
8” TFT screen with friendly user interface
-
Automatic or manual testing of basic error, creeping, starting, deviation
-
Online diagnostics and self-protection for overload, short-circuit, or open-circuit
-
Harmonic output up to 63rd order
-
Supports DLT-645 communication protocol for multifunction meters
-
Stand-alone, PC, or remote control via RS232, Ethernet, WiFi, RS485
-
Auto-calibration schemes with user-defined testing sequences
-
Portable yet powerful system (35 kg, compact housing)
Technical Specifications
AC Voltage Output
-
Range: 100V, 220V, 380V
-
Adjusting Range: 0–120% RG
-
Resolution: 0.01%RG, 0.1%RG, 1%RG, or 10%RG
-
Stability: 0.01%/2min (Class 0.05), 0.02%/2min (Class 0.1)
-
Distortion: ≤0.2% (Non-capacitive load)
-
Max Output Load: 20 VA/phase
-
Accuracy: Class 0.05% or 0.1%
AC Current Output
-
Range: 0.05A, 0.2A, 1A, 5A, 20A, 100A
-
Adjusting Range: 0–120% RG
-
Resolution: 0.01%RG, 0.1%RG, 1%RG, 10%RG
-
Stability: 0.01%/2min (Class 0.05), 0.02%/2min (Class 0.1)
-
Distortion: ≤0.2% (Non-capacitive load)
-
Max Output Load: 120 VA (100A range)
-
Accuracy: Class 0.05% or 0.1%
Power Output
-
Stability: 0.01%/2min (Class 0.05), 0.02%/2min (Class 0.1)
-
Active/Reactive Power Accuracy: 0.05% RG
Phase Output
-
Range: 0°–360°
-
Resolution: 0.01° (steps of 10°, 1°, 0.1°, or 0.01°)
-
Accuracy: ±0.05°
Power Factor Output
-
Range: -1 to +1
-
Resolution: 0.0001
-
Accuracy: 0.0005
Frequency Output
-
Range: 40Hz–70Hz
-
Resolution: 0.001Hz (steps of 1Hz, 0.1Hz, 0.01Hz, or 0.001Hz)
-
Accuracy: ±0.002Hz
Harmonic Setting
-
Order: 2nd–63rd
-
Amplitude: 0–40%
-
Angle: 0°–359.99°
-
Error: ≤±0.1% (2nd–31st), ≤±0.2% (32nd–63rd)
Energy Error Measurement
-
Active Energy Basic Error: ±0.05% RD or ±0.1% RD (depending on class)
-
Reactive Energy Basic Error: ±0.1% RD or ±0.2% RD (depending on class)
-
Input Pulse: 3 channels active, 3 channels reactive
Other Parameters
-
Power Supply: 90–265 VAC/DC, 50–60Hz
-
Power Consumption: 50–1000 VA
-
Operating Temp: 20°C–30°C, RH ≤85%
-
Storage Temp: -20°C–50°C
-
Dimensions: 600 × 440 × 266 mm
-
Weight: 35 kg
Main Functions
-
Calibration of single/three-phase electronic or inductive meters
-
Automatic/manual testing of basic error, creeping, starting, deviation
-
Variable influence tests: voltage, frequency, harmonics, reverse phase sequence, voltage unbalance
-
Special energy meter type tests: phase triggering, pulse train triggering
-
Multifunction meter tests: demand indication error, demand cycle error, time slot switching error, multi-tariff energy error
-
With OCXO, capable of detecting daily error of energy meters
.png)