A‑Series OLT Oil Breakdown Voltage Tester
The OLT A-Series oil testers are compact, lightweight, and fully automatic systems designed for dielectric breakdown voltage testing of insulating liquids.
Built in compliance with IEC 60156 and safety standards IEC 61010 / IEC 61326-1.
Preloaded with international test sequences (IEC, GOST, ASTM) for immediate use.
Features a built-in printer, internal data storage, and PC connectivity for reporting.
Measures voltage directly on electrodes with a true sine HV output, ensuring accuracy and repeatability.
Includes an integrated oil temperature measurement function.
Supplied ready to operate, complete with IEC 60156 test vessel, VDE 0370 mushroom electrodes, Go/No-Go gauge, and accessories.
Key Features
-
Compact and lightweight
-
Fully automatic operation
-
Fully compatible with IEC 60156
-
Preloaded programs for IEC, GOST, ASTM standards
-
Built-in thermal printer
-
Internal memory (512 results)
-
PC connectivity via USB
-
True sine high-voltage on electrodes
-
Direct voltage measurement
-
Integrated oil temperature measurement
-
Safety interlocks with automatic HV cutoff
Technical Specifications
output ac voltage (symmetrical)
-
80 / 90 / 100 kV rms (OLT-80A / OLT-90A / OLT-100A)
voltage resolution
-
0.1 kV
measurement uncertainty
-
±3% of reading
output voltage rise rate
-
2 ± 0.2 kV/s (programmable per standard)
switch-off time on breakdown
-
< 5 µs
temperature measuring range
-
5.0 – 50.0 °C
programmed test sequences
-
IEC 60156, GOST 6581, ASTM D877, ASTM D1816 (optional)
electrodes
-
Mushroom type (IEC 60156 basic)
-
ASTM D877 / ASTM D1816 (optional)
test vessel
-
IEC 60156, 400 ml, 1.0 kg
go/no-go gauge width
-
2.45 / 2.55 mm (basic)
-
Optional: 1, 2, 2.5, 4 mm
display resolution
-
128 × 64 dot matrix, monochrome
interface languages
-
English, Russian, Turkish
printer
-
Thermal line, 8 dots/mm, 384 dots/line
memory for results
-
512 test records
pc connectivity
-
USB B-type
input voltage
-
230 V ±10% (standard)
-
110/120 V available on request
line frequency
-
50 / 60 Hz ±1
power consumption (max)
-
OLT-80A: 100 VA
-
OLT-90A: 120 VA
-
OLT-100A: 140 VA
dimensions
-
OLT-80A: 461 × 280 × 271 mm
-
OLT-90A: 420 × 300 × 320 mm
-
OLT-100A: 530 × 315 × 320 mm
net weight
-
OLT-80A: 21 kg
-
OLT-90A: 25 kg
-
OLT-100A: 30 kg
environment
-
Operating temperature: +10 … +55 °C
-
Storage temperature: +5 … +60 °C
-
Relative humidity: ≤80% RH @ 25 °C (non-condensing)
-
Atmospheric pressure: 630 – 800 mmHg
Application
-
Transformer oil dielectric strength testing
-
Acceptance and routine testing in utilities and transformer workshops
-
R&D and laboratory use for quality control of insulating liquids
Standard Accessories
-
OLT A-Series Tester (80A / 90A / 100A)
-
IEC 60156 test vessel with electrodes and cover
-
Go/No-Go gauge
-
Magnetic stirrer + lifting stick
-
USB cable (A-B type)
-
Power cord (C-type plug)
-
Grounding cable
-
Software on USB flash drive
-
Thermal printer paper roll (installed)
-
Fuses (250V/3A) × 2
-
User’s manual
-
Optional ISO/IEC 17025 calibration certificate
-
Optional calibration kit C-80 / C-100
.png)