PDGA.100 Portable Dissolved Gas Analyser
About the PDGA.100:
The sample flows into the vaporizer and is directed to two chromatographic columns:
-
Column 1 (GDX502): Used to separate hydrocarbon compound components, with examination conducted on FID1.
-
Column 2 (TDX-01): Used to separate H₂, CO, and CO₂. The sample first passes through the TCD for H₂ component examination. After the TCD, CO and CO₂ are transformed into CH₄ via a re-burner, and the examination is conducted on FID2.
This system is compliant with IEC-60567-2011 and IEC-60599-1999 standards.
Standard Accessories
-
Gas Extraction Syringe
-
3-Way-Valve
-
Plunger Rubber Head
-
Rubber Cap
-
Injection Septum
-
5mL Glass Syringe for gas collection
-
1mL Glass Syringe for sample injection
-
200mL Sampling Container with PU Pipe
-
Nut for Gas Circuit
-
Silicon Rubber Sealing Ring for Gas Circuit
-
WH500-USB Workstation Installation Disk (Specialized for electric system)
-
USB Cable
-
Industrial Transport Case IP67 (When Case Closed)
Specifications:
Temperature Control Range
-
Oven: RT + 6°C to 400°C (1°C increment)
-
FID, TCD, Injector, Re-burner: RT + 8°C to 400°C
LCD Screen
-
7-inch screen
PC Interface
-
USB
Temperature Control Accuracy
-
TCD: ±0.02°C
-
Oven, FID, Injector, Re-burner: ±0.1°C
FID Limit of Detection
-
≤2×10⁻¹² g/s (N-C16)
TCD Sensitivity
-
≥8000 mV·mL/mg (N-C16)
Minimum Detection Concentration
-
H₂: ≤2 ppm
-
CO/CO₂: ≤2 ppm
-
CH₄/C₂H₂/C₂H₄/C₂H₆: ≤0.1 ppm
Oil Types
-
Synthetic & Natural Oils
Sample Volume
-
5 mL – 50 mL
Operation Pressure
-
0.08 MPa – 0.3 MPa (800 – 3000 millibar)
Operation Temperature
-
0°C - 65°C
Input Power
-
220 VAC / 50 Hz
Dimensions
-
380 mm × 500 mm × 220 mm (L × W × H)
-
Microprocessor-controlled automatic withstand voltage test for oil circulation (3 x 0 ~ 100KV), including boosting, maintaining, mixing, standing, printing, and data export
-
Large LCD display with touch screen operation
-
User-friendly, fully automated withstand voltage testing on one sample
-
Breakdown voltage value recorded 1 to 9 times and automatically saved
-
Thermal printer prints breakdown voltage values and average value after the test
-
Power-down data preservation
-
Stores up to 100 test results
-
Accurate voltage at 50Hz
-
Over-voltage and over-current protection for operator safety
-
Displays time & date on test results
-
RS232 & USB interface for data export