top of page
PDGA.100 Portable Dissolved Gas Analyser
About the PDGA.100:

The sample flows into the vaporizer and is directed to two chromatographic columns:

  • Column 1 (GDX502): Used to separate hydrocarbon compound components, with examination conducted on FID1.

  • Column 2 (TDX-01): Used to separate H₂, CO, and CO₂. The sample first passes through the TCD for H₂ component examination. After the TCD, CO and CO₂ are transformed into CH₄ via a re-burner, and the examination is conducted on FID2.

This system is compliant with IEC-60567-2011 and IEC-60599-1999 standards.

Standard Accessories
  • Gas Extraction Syringe

  • 3-Way-Valve

  • Plunger Rubber Head

  • Rubber Cap

  • Injection Septum

  • 5mL Glass Syringe for gas collection

  • 1mL Glass Syringe for sample injection

  • 200mL Sampling Container with PU Pipe

  • Nut for Gas Circuit

  • Silicon Rubber Sealing Ring for Gas Circuit

  • WH500-USB Workstation Installation Disk (Specialized for electric system)

  • USB Cable

  • Industrial Transport Case IP67 (When Case Closed)

Specifications:

Temperature Control Range

  • Oven: RT + 6°C to 400°C (1°C increment)

  • FID, TCD, Injector, Re-burner: RT + 8°C to 400°C

LCD Screen

  • 7-inch screen

PC Interface

  • USB

Temperature Control Accuracy

  • TCD: ±0.02°C

  • Oven, FID, Injector, Re-burner: ±0.1°C

FID Limit of Detection

  • ≤2×10⁻¹² g/s (N-C16)

TCD Sensitivity

  • ≥8000 mV·mL/mg (N-C16)

Minimum Detection Concentration

  • H₂: ≤2 ppm

  • CO/CO₂: ≤2 ppm

  • CH₄/C₂H₂/C₂H₄/C₂H₆: ≤0.1 ppm

Oil Types

  • Synthetic & Natural Oils

Sample Volume

  • 5 mL – 50 mL

Operation Pressure

  • 0.08 MPa – 0.3 MPa (800 – 3000 millibar)

Operation Temperature

  • 0°C - 65°C

Input Power

  • 220 VAC / 50 Hz

Dimensions

  • 380 mm × 500 mm × 220 mm (L × W × H)

  • Microprocessor-controlled automatic withstand voltage test for oil circulation (3 x 0 ~ 100KV), including boosting, maintaining, mixing, standing, printing, and data export

  • Large LCD display with touch screen operation

  • User-friendly, fully automated withstand voltage testing on one sample

  • Breakdown voltage value recorded 1 to 9 times and automatically saved

  • Thermal printer prints breakdown voltage values and average value after the test

  • Power-down data preservation

  • Stores up to 100 test results

  • Accurate voltage at 50Hz

  • Over-voltage and over-current protection for operator safety

  • Displays time & date on test results

  • RS232 & USB interface for data export

bottom of page